dc.description.abstractPurpose Computed tomography (CT) and, in particular, cone beam CT (CBCT) have been increasingly used as a diagnostic tool in recent years. Patient motion during acquisition is common in CBCT due to long scan times. This results in degraded image quality and may potentially increase the number of retakes. Our aim was to develop a marker-free iterative motion correction algorithm that works on the projection images and is suitable for local tomography. Methods We present an iterative motion correction algorithm that allows the patient's motion to be detected and taken into account during reconstruction. The core of our method is a fast GPU-accelerated three-dimensional reconstruction algorithm. Assuming rigid motion, motion correction is performed by minimizing a pixel-wise cost function between all captured x-ray images and parameterized projections of the reconstructed volume. Results Our method is marker-free and requires only projection images. Furthermore, it can deal with local tomography data. We demonstrate the effectiveness of our approach on both simulated and real motion-beset patient images. The results show that our new motion correction algorithm leads to accurate reconstructions with sharper edges, better contrasts and more detail. Conclusions The presented method allows for correction of patient motion with observable improvements in image quality compared to uncorrected reconstructions. Potentially, this may reduce the number of retakes caused by corrupted reconstructions due to patient movements.en_GB
dc.language.isoengde
dc.rightsCC BY*
dc.rights.urihttps://creativecommons.org/licenses/by/4.0/*
dc.subject.ddc004 Informatikde_DE
dc.subject.ddc004 Data processingen_GB
dc.titleProjection-based improvement of 3D reconstructions from motion-impaired dental cone beam CT dataen_GB
dc.typeZeitschriftenaufsatzde
dc.identifier.doihttp://doi.org/10.25358/openscience-7038-
jgu.type.contenttypeScientific articlede
jgu.type.dinitypearticleen_GB
jgu.type.versionPublished versionde
jgu.type.resourceTextde
jgu.organisation.departmentFB 08 Physik, Mathematik u. Informatikde
jgu.organisation.number7940-
jgu.organisation.nameJohannes Gutenberg-Universität Mainz-
jgu.rights.accessrightsopenAccess-
jgu.journal.titleMedical physicsde
jgu.journal.volume46de
jgu.journal.issue10de
jgu.pages.start4470de
jgu.pages.end4480de
jgu.publisher.year2019-
jgu.publisher.nameWileyde
jgu.publisher.placeHoboken, NJde
jgu.publisher.issn2473-4209de
jgu.publisher.issn1522-8541de
jgu.publisher.issn0094-2405de
jgu.organisation.placeMainz-
jgu.subject.ddccode004de
jgu.publisher.doi10.1002/mp.13731de
jgu.organisation.rorhttps://ror.org/023b0x485