Projection-based improvement of 3D reconstructions from motion-impaired dental cone beam CT data

dc.contributor.authorNiebler, Stefan
dc.contributor.authorSchömer, Elmar
dc.contributor.authorTjaden, Henning
dc.contributor.authorSchwanecke, Ulrich
dc.contributor.authorSchulze, Ralf
dc.date.accessioned2022-05-25T09:41:44Z
dc.date.available2022-05-25T09:41:44Z
dc.date.issued2019
dc.description.abstractPurpose Computed tomography (CT) and, in particular, cone beam CT (CBCT) have been increasingly used as a diagnostic tool in recent years. Patient motion during acquisition is common in CBCT due to long scan times. This results in degraded image quality and may potentially increase the number of retakes. Our aim was to develop a marker-free iterative motion correction algorithm that works on the projection images and is suitable for local tomography. Methods We present an iterative motion correction algorithm that allows the patient's motion to be detected and taken into account during reconstruction. The core of our method is a fast GPU-accelerated three-dimensional reconstruction algorithm. Assuming rigid motion, motion correction is performed by minimizing a pixel-wise cost function between all captured x-ray images and parameterized projections of the reconstructed volume. Results Our method is marker-free and requires only projection images. Furthermore, it can deal with local tomography data. We demonstrate the effectiveness of our approach on both simulated and real motion-beset patient images. The results show that our new motion correction algorithm leads to accurate reconstructions with sharper edges, better contrasts and more detail. Conclusions The presented method allows for correction of patient motion with observable improvements in image quality compared to uncorrected reconstructions. Potentially, this may reduce the number of retakes caused by corrupted reconstructions due to patient movements.en_GB
dc.identifier.doihttp://doi.org/10.25358/openscience-7038
dc.identifier.urihttps://openscience.ub.uni-mainz.de/handle/20.500.12030/7052
dc.language.isoengde
dc.rightsCC-BY-4.0*
dc.rights.urihttps://creativecommons.org/licenses/by/4.0/*
dc.subject.ddc004 Informatikde_DE
dc.subject.ddc004 Data processingen_GB
dc.titleProjection-based improvement of 3D reconstructions from motion-impaired dental cone beam CT dataen_GB
dc.typeZeitschriftenaufsatzde
jgu.journal.issue10de
jgu.journal.titleMedical physicsde
jgu.journal.volume46de
jgu.organisation.departmentFB 08 Physik, Mathematik u. Informatikde
jgu.organisation.nameJohannes Gutenberg-Universität Mainz
jgu.organisation.number7940
jgu.organisation.placeMainz
jgu.organisation.rorhttps://ror.org/023b0x485
jgu.pages.end4480de
jgu.pages.start4470de
jgu.publisher.doi10.1002/mp.13731de
jgu.publisher.issn2473-4209de
jgu.publisher.issn1522-8541de
jgu.publisher.issn0094-2405de
jgu.publisher.nameWileyde
jgu.publisher.placeHoboken, NJde
jgu.publisher.year2019
jgu.rights.accessrightsopenAccess
jgu.subject.ddccode004de
jgu.type.contenttypeScientific articlede
jgu.type.dinitypeArticleen_GB
jgu.type.resourceTextde
jgu.type.versionPublished versionde

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
projectionbased_improvement_o-20220525114531380.pdf
Size:
581.38 KB
Format:
Adobe Portable Document Format

License bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
3.57 KB
Format:
Item-specific license agreed upon to submission
Description: