Variability of transient elastography-based spleen stiffness performed at 100 Hz
Loading...
Date issued
Editors
Journal Title
Journal ISSN
Volume Title
Publisher
Reuse License
Description of rights: CC-BY-4.0
Abstract
Background
Spleen stiffness measurement (SSM) performed by transient elastography at 100 Hz is a novel technology for the evaluation of portal hypertension in advanced chronic liver disease, but technical aspects are lacking. We aimed to evaluate the intraexamination variability of SSM and to determine the best transient elastography protocol for obtaining robust measurements to be used in clinical practice.
Methods
We analyzed 253 SSM exams with up to 20 scans for each examination, performed between April 2021 and June 2022. All SSM results were evaluated according to different protocols by dividing data into groups of n measurements (from 2 to 19). Considering as reference the median SSM values across all the 20 measurements, we calculated the distribution of the absolute deviations of each protocol from the reference median. This analysis was repeated 1,000 times by resampling the data. Distributions were also stratified by etiology (chronic liver disease versus clinically significant portal hypertension) and different SSM ranges: < 25 kPa, 25–75, and > 75 kPa.
Results
Overall, we observed that the spleen stiffness exam had less variability if it exceeded 12 measurements, i.e., absolute deviations ≤ 5 kPa at 95% confidence. For exams with higher SSM values (> 75 kPa), as seen in clinically significant portal hypertension, at least 15 measurements are highly recommendable.
Conclusions
Fifteen scans per examination should be considered for each SSM exam performed at 100 Hz to achieve a low intraexamination variability within a reasonable time in clinical practice.
Description
Keywords
Citation
Published in
European radiology experimental, 7, Springer International Publishing, Dordrecht u.a., 2023, https://doi.org/10.1186/s41747-023-00393-2
