A rapid point-of-care assay accurately measures vitamin D
Date issued
Editors
Journal Title
Journal ISSN
Volume Title
Publisher
License
Abstract
Purpose
Vitamin D (VitD) is a pleiotropic hormone with effects on a multitude of systems and metabolic pathways. Consequently, the relevance of a sufficiently high VitD serum level becomes self-evident.
Methods
A rapid immunofluorescence assay designed for the point-of-care measurement of serum VitD3 solely was tested. Inter- and intra-assay validation, double testing and result comparison with a standardized laboratory method were performed.
Results
An overall linear correlation of r = 0.89 (Pearson, 95% CI 0.88–0.92, p < 0.01) between the point of care and the conventional reference assay was registered. Accuracy and precision were of special interest at cut-points (10 ng/ml [mean deviation 1.7 ng/ml, SD 1.98 ng/ml, SE 0.16 ng/ml], 12 ng/ml [MD 0.41, SD 1.89, SE 0.19] and 30 ng/ml [MD − 1.11, SD 3.89, SE 0.35]). Only a slight deviation was detected between the two assays when using fresh (r = 0.91, 95% CI 0.86–0.94, p < 0.01) and frozen serum samples (r = 0.86, 0.82–0.89, p < 0.01). Results remained steady when samples were frozen several times. Inter- and intra-assay validation according to the CLSI protocol as well as multiuser testing showed stable results.
Conclusion
This novel, innovative, and controlled study indicates that the evaluated rapid point of care VitD assay is reliable, accurate, and suited for clinical practice.
Description
Keywords
Citation
Published in
Journal of endocrinological investigation, 44, Springer, S. l., 2021, https://doi.org/10.1007/s40618-021-01575-8