Autoren: | Wittenmeier, Eva Paumen, Yuri Mildenberger, Philipp Smetiprach, Julia Pirlich, Nina Griemert, Eva-Verena Kriege, Marc Engelhard, Kristin |
Titel: | Non-invasive haemoglobin measurement as an index test to detect pre-operative anaemia in elective surgery patients : a prospective study |
Online-Publikationsdatum: | 27-Apr-2022 |
Erscheinungsdatum: | 2021 |
Sprache des Dokuments: | Englisch |
Zusammenfassung/Abstract: | Non-invasive haemoglobin measurement using absolute values lacks the precision to be the sole basis for the treatment of pre-operative anaemia. However, it can possibly serve as a screening test, indexing ‘anaemia’ with high sensitivity when values remain under prespecified cut-off values. Based on previous data, non-invasive haemoglobin cut-off values (146 g.l−1 for women and 152 g.l−1 for men) detect true anaemia with 99% sensitivity. An index test with these prespecified cut-off values was verified by prospective measurement of non-invasive and invasive haemoglobin pre-operatively in elective surgical patients. In 809 patients, this showed an estimated sensitivity (95%CI) of 98.9% (94.1–99.9%) in women and 96.4% (91.0–99.0%) in men. This saved invasive blood tests in 9% of female and 28% of male patients. In female patients, a lower non-invasive haemoglobin cut-off value (138 g.l−1) would save 28% of invasive blood tests with a sensitivity of 95%. The target 99% sensitivity would be reached by non-invasive haemoglobin cut-off values of 152 g.l−1 in female and 162 g.l−1 in male patients, saving 3% and 9% of invasive blood tests, respectively. Bias and limits of agreement between non-invasive and laboratory haemoglobin levels were 2 and − 25 to 28 g.l−1, respectively. Patient and measurement characteristics did not influence the agreement between non-invasive and laboratory haemoglobin levels. Although sensitivity was very high, the index test using prespecified cut-off values just failed to reach the target sensitivity to detect true anaemia. Nevertheless, with respect to blood-sparing effects, the use of the index test in men may be clinically useful, while an index test with a lower cut-off (132 g.l−1) could be more clinically appropriate in women. |
DDC-Sachgruppe: | 610 Medizin 610 Medical sciences |
Veröffentlichende Institution: | Johannes Gutenberg-Universität Mainz |
Organisationseinheit: | FB 04 Medizin |
Veröffentlichungsort: | Mainz |
ROR: | https://ror.org/023b0x485 |
DOI: | http://doi.org/10.25358/openscience-6910 |
Version: | Published version |
Publikationstyp: | Zeitschriftenaufsatz |
Nutzungsrechte: | CC BY-NC |
Informationen zu den Nutzungsrechten: | https://creativecommons.org/licenses/by-nc/4.0/ |
Zeitschrift: | Anaesthesia 76 5 |
Seitenzahl oder Artikelnummer: | 647 654 |
Verlag: | Wiley-Blackwell |
Verlagsort: | Oxford u.a |
Erscheinungsdatum: | 2021 |
ISSN: | 1365-2044 |
DOI der Originalveröffentlichung: | 10.1111/anae.15312 |