Prospective comparison of handheld ultrasound devices from different manufacturers with respect to B-scan quality and clinical significance for various abdominal sonography questions

dc.contributor.authorMerkel, Daniel
dc.contributor.authorZüllich, Tim Felix
dc.contributor.authorSchneider, Christoph
dc.contributor.authorYousefzada, Masuod
dc.contributor.authorBeer, Diana
dc.contributor.authorLudwig, Michael
dc.contributor.authorWeimer, Andreas
dc.contributor.authorKünzel, Julian
dc.contributor.authorKloeckner, Roman
dc.contributor.authorWeimer, Johannes Matthias
dc.date.accessioned2024-02-20T09:37:38Z
dc.date.available2024-02-20T09:37:38Z
dc.date.issued2023
dc.description.abstractBackground: Handheld ultrasound (HHUS) devices have chiefly been deployed in emergency medicine, where they are considered a valid tool. The data situation is less clear in the case of internal questions in abdominal sonography. In our study, we investigate whether HHUS devices from different manufacturers differ in their B-scan quality, and whether any differences are relevant for the significance of an internal ultrasound examination. Method: The study incorporated eight HHUS devices from different manufacturers. Ultrasound videos of seven defined sonographic questions were recorded with all of the devices. The analogue recording of the same findings with a conventional high-end ultrasound (HEUS) device served as an evaluation criterion. Then, the corresponding findings were played side by side and evaluated by fourteen ultrasound experts using a point scale (5 points = very good; 1 point = insufficient). Results: The HHUS devices achieved relatively good results in terms of both the B-scan quality assessment and the ability to answer the clinical question, regardless of the manufacturer. One of the tested HHUS devices even achieved a significantly (p < 0.05) higher average points score in both the evaluation of B-scan quality and in the evaluation of clinical significance than the other devices. Regardless of the manufacturer, the HHUS devices performed best when determining the status/inferior vena cava volume and in the representation of ascites/free fluid. Conclusion: In various clinical abdominal sonography questions, HHUS systems can reliably reproduce findings, and are—while bearing their limitations in mind—an acceptable alternative to conventional HEUS systems. Irrespective of this, the present study demonstrated relevant differences in the B-scan quality of HHUS devices from different manufacturers.en_GB
dc.identifier.doihttp://doi.org/10.25358/openscience-10083
dc.identifier.urihttps://openscience.ub.uni-mainz.de/handle/20.500.12030/10101
dc.language.isoeng
dc.rightsCC-BY-4.0
dc.rights.urihttps://creativecommons.org/licenses/by/4.0/
dc.subject.ddc610 Medizinde_DE
dc.subject.ddc610 Medical sciencesen_GB
dc.titleProspective comparison of handheld ultrasound devices from different manufacturers with respect to B-scan quality and clinical significance for various abdominal sonography questionsen_GB
dc.typeZeitschriftenaufsatzde_DE
jgu.apc.netprice2476,02
jgu.apc.price2649,34
jgu.apc.taxrate7
jgu.dfg.year2023
jgu.journal.issue24
jgu.journal.titleDiagnostics
jgu.journal.volume13
jgu.nationalcurrency.eur2476,02
jgu.organisation.departmentFB 04 Medizinde_DE
jgu.organisation.nameJohannes Gutenberg-Universität Mainzde_DE
jgu.organisation.number2700
jgu.organisation.placeMainz
jgu.organisation.rorhttps://ror.org/023b0x485
jgu.pages.alternative3622
jgu.publisher.doi10.3390/diagnostics13243622
jgu.publisher.issn2075-4418
jgu.publisher.nameMDPI
jgu.publisher.placeBasel
jgu.publisher.year2023
jgu.rights.accessrightsopenAccessen_GB
jgu.subject.ddccode610
jgu.subject.dfgMultidisciplinaryde_DE
jgu.type.contenttypeScientific articleen_GB
jgu.type.dinitypeArticleen_GB
jgu.type.resourceTexten_GB
jgu.type.versionPublished versionen_GB

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
prospective_comparison_of_han-20240214144126010.pdf
Size:
2.56 MB
Format:
Adobe Portable Document Format
Description:

License bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
3.57 KB
Format:
Item-specific license agreed upon to submission
Description:

Collections