Titel: Impact of different screw designs on durability of fracture fixation : in vitro study with cyclic loading of scaphoid bones
Online-Publikationsdatum: 5-Okt-2022
Erscheinungsdatum: 2016
Sprache des Dokuments: Englisch
Zusammenfassung/Abstract: Purpose The use of new headless compression screws (HCSs) for scaphoid fixation is growing, but the nonunion rate has remained constant. The aim of this study was to compare the stability of fixation resulting from four modern HCSs using a simulated fracture model to determine the optimal screw design(s). Methods We tested 40 fresh-frozen cadaver scaphoids treated with the Acumed Acutrak 2 mini (AA), the KLS Martin HBS2 midi (MH), the Stryker TwinFix (ST) and the Synthes HCS 3.0 with a long thread (SH). The bones with simulated fractures and implanted screws were loaded uniaxially into flexion for 2000 cycles with a constant bending moment of 800 Nmm. The angulation of the fracture fragments was measured continuously. Data were assessed statistically using the univariate ANOVA test and linear regression analysis, and the significance level was set at p < 0.05. Results The median angulation of bone fragments φ allowed by each screw was 0.89° for AA, 1.12° for ST, 1.44° for SH and 2.36° for MH. With regards to linear regression, the most reliable curve was achieved by MH, with a coefficient of determination of R2 = 0.827. This was followed by AA (R2 = 0.354), SH (R2 = 0.247) and ST (R2 = 0.019). Data assessed using an adapted ANOVA model showed no statistically significant difference (p = 0.291) between the screws. Conclusions The continuous development of HCSs has resulted in very comparable implants, and thus, at this time, other factors, such as surgeons’ experience, ease of handling and price, should be taken into consideration.
DDC-Sachgruppe: 610 Medizin
610 Medical sciences
Veröffentlichende Institution: Johannes Gutenberg-Universität Mainz
Organisationseinheit: FB 04 Medizin
Veröffentlichungsort: Mainz
ROR: https://ror.org/023b0x485
DOI: http://doi.org/10.25358/openscience-7847
Version: Published version
Publikationstyp: Zeitschriftenaufsatz
Nutzungsrechte: CC BY
Informationen zu den Nutzungsrechten: https://creativecommons.org/licenses/by/4.0/
Zeitschrift: PLoS one
11
1
Seitenzahl oder Artikelnummer: e0145949
Verlag: PLoS
Verlagsort: Lawrence, Kan.
Erscheinungsdatum: 2016
ISSN: 1932-6203
URL der Originalveröffentlichung: http://dx.doi.org/10.1371/journal.pone.0145949
DOI der Originalveröffentlichung: 10.1371/journal.pone.0145949
Enthalten in den Sammlungen:DFG-OA-Publizieren (2012 - 2017)

Dateien zu dieser Ressource:
  Datei Beschreibung GrößeFormat
Miniaturbild
impact_of_different_screw_des-20220914005604819.pdf1.62 MBAdobe PDFÖffnen/Anzeigen