Intra-individual comparison of epicardial adipose tissue characteristics on coronary CT angiography between photon-counting detector and energy-integrating detector CT systems
Date issued
Editors
Journal Title
Journal ISSN
Volume Title
Publisher
License
Abstract
Purpose
To explore the potential differences in epicardial adipose tissue (EAT) volume and attenuation measurements between photon-counting detector (PCD) and energy-integrating detector (EID)-CT systems.
Methods
Fifty patients (mean age 69 ± 8 years, 41 male [82 %]) were prospectively enrolled for a research coronary CT angiography (CCTA) on a PCD-CT within 30 days after clinical EID-based CCTA. EID-CT acquisitions were reconstructed using a Bv40 kernel at 0.6 mm slice thickness. The PCD-CT acquisition was reconstructed at a down-sampled resolution (0.6 mm, Bv40; [PCD-DS]) and at ultra-high resolutions (PCD-UHR) with a 0.2 mm slice thickness and Bv40, Bv48, and Bv64 kernels. EAT segmentation was performed semi-automatically at about 1 cm intervals and interpolated to cover the whole epicardium within a threshold of −190 to −30 HU. A subgroup analysis was performed based on quartile groups created from EID-CT data and PCD-UHRBv48 data. Differences were measured using repeated-measures ANOVA and the Friedman test. Correlations were tested using Pearson’s and Spearman’s rho, and agreement using Bland-Altman plots.
Results
EAT volumes significantly differed between some reconstructions (e.g. EID-CT: 138 ml [IQR 100, 188]; PCD-DS: 147 ml [110, 206]; P<0.001). Overall, correlations between PCD-UHR and EID-CT EAT volumes were excellent, e.g. PCD-UHRBv48: r: 0.976 (95 % CI: 0.958, 0.987); P<0.001; with good agreement (mean bias: −9.5 ml; limits of agreement [LoA]: −40.6, 21.6). On the other hand, correlations regarding EAT attenuation was moderate, e.g. PCD-UHRBV48: r: 0.655 (95 % CI: 0.461, 0.790); P<0.001; mean bias: 6.5 HU; LoA: −2.0, 15.0.
Conclusion
EAT attenuation and volume measurements demonstrated different absolute values between PCD-UHR, PCD-DS as well as EID-CT reconstructions, but showed similar tendencies on an intra-individual level. New protocols and threshold ranges need to be developed to allow comparison between PCD-CT and EID-CT data.