Motion correction for separate mandibular and cranial movements in cone beam CT reconstructions

dc.contributor.authorBirklein, Lukas
dc.contributor.authorNiebler, Stefan
dc.contributor.authorSchömer, Elmar
dc.contributor.authorBrylka, Robert
dc.contributor.authorSchwanecke, Ulrich
dc.contributor.authorSchulze, Ralf
dc.date.accessioned2023-08-03T07:44:20Z
dc.date.available2023-08-03T07:44:20Z
dc.date.issued2023
dc.date.updated2023-07-24T12:09:44Z
dc.description.abstractBackground Patient motions are a repeatedly reported phenomenon in oral and maxillofacial cone beam CT scans, leading to reconstructions of limited usability. In certain cases, independent movements of the mandible induce unpredictable motion patterns. Previous motion correction methods are not able to handle such complex cases of patient movements. Purpose Our goal was to design a combined motion estimation and motion correction approach for separate cranial and mandibular motions, solely based on the 2D projection images from a single scan. Methods Our iterative three-step motion correction algorithm models the two articulated motions as independent rigid motions. First of all, we segment cranium and mandible in the projection images using a deep neural network. Next, we compute a 3D reconstruction with the poses of the object's trajectories fixed. Third, we improve all poses by minimizing the projection error while keeping the reconstruction fixed. Step two and three are repeated alternately. Results We find that our marker-free approach delivers reconstructions of up to 85% higher quality, with respect to the projection error, and can improve on already existing techniques, which model only a single rigid motion. We show results of both synthetic and real data created in different scenarios. The reconstruction of motion parameters in a real environment was evaluated on acquisitions of a skull mounted on a hexapod, creating a realistic, easily reproducible motion profile. Conclusions The proposed algorithm consistently enhances the visual quality of motion impaired cone beam computed tomography scans, thus eliminating the need for a re-scan in certain cases, considerably lowering radiation dosage for the patient. It can flexibly be used with differently sized regions of interest and is even applicable to local tomography.en_GB
dc.description.sponsorshipDeutsche Forschungsgemeinschaft (DFG)|491381577|Open-Access-Publikationskosten 2022–2024 Universität Mainz - Universitätsmedizin
dc.identifier.doihttp://doi.org/10.25358/openscience-9342
dc.identifier.urihttps://openscience.ub.uni-mainz.de/handle/20.500.12030/9360
dc.language.isoengde
dc.rightsCC-BY-4.0*
dc.rights.urihttps://creativecommons.org/licenses/by/4.0/*
dc.subject.ddc004 Informatikde_DE
dc.subject.ddc004 Data processingen_GB
dc.titleMotion correction for separate mandibular and cranial movements in cone beam CT reconstructionsen_GB
dc.typeZeitschriftenaufsatzde
elements.object.id154581
elements.object.labelscone beam computed tomography
elements.object.labelsmotion
elements.object.labelstomography
elements.object.labelsSkull
elements.object.labelsMandible
elements.object.labelsHumans
elements.object.labelsArtifacts
elements.object.labelsPhantoms, Imaging
elements.object.labelsMovement
elements.object.labelsAlgorithms
elements.object.labelsMotion
elements.object.labelsImage Processing, Computer-Assisted
elements.object.labelsCone-Beam Computed Tomography
elements.object.labelscone beam computed tomography
elements.object.labelsmotion
elements.object.labelstomography
elements.object.labelsHumans
elements.object.labelsMovement
elements.object.labelsMotion
elements.object.labelsCone-Beam Computed Tomography
elements.object.labelsSkull
elements.object.labelsMandible
elements.object.labelsAlgorithms
elements.object.labelsImage Processing, Computer-Assisted
elements.object.labelsPhantoms, Imaging
elements.object.labelsArtifacts
elements.object.labels0299 Other Physical Sciences
elements.object.labels0903 Biomedical Engineering
elements.object.labels1112 Oncology and Carcinogenesis
elements.object.labelsNuclear Medicine & Medical Imaging
elements.object.labels4003 Biomedical engineering
elements.object.labels5105 Medical and biological physics
elements.object.typejournal-article
jgu.journal.issue6de
jgu.journal.titleMedical physicsde
jgu.journal.volume50de
jgu.organisation.departmentFB 08 Physik, Mathematik u. Informatikde
jgu.organisation.nameJohannes Gutenberg-Universität Mainz
jgu.organisation.number7940
jgu.organisation.placeMainz
jgu.organisation.rorhttps://ror.org/023b0x485
jgu.pages.end3525de
jgu.pages.start3511de
jgu.publisher.doi10.1002/mp.16347de
jgu.publisher.issn0094-2405de
jgu.publisher.licenceCC BY
jgu.publisher.nameWileyde
jgu.publisher.placeHoboken, NJde
jgu.publisher.year2023
jgu.rights.accessrightsopenAccess
jgu.subject.ddccode004de
jgu.subject.dfgNaturwissenschaftende
jgu.type.dinitypeArticleen_GB
jgu.type.resourceTextde
jgu.type.versionPublished versionde

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
motion_correction_for_separat-20230724140945582.pdf
Size:
3.12 MB
Format:
Adobe Portable Document Format
Description:
Published version

Collections