Abstract
Purpose
We evaluated a dedicated dose-reduced UHR-CT for head and neck imaging, combined with a novel deep learning reconstruction algorithm to assess its impact on image quality and radiation exposure.
Methods
Retrospective analysis of ninety-eight consecutive patients examined using a new body weight-adapted protocol. Images were reconstructed using adaptive iterative dose reduction and advanced intelligent Clear-IQ engine with an already established (DL-1) and a newly implemented reconstruction algorithm (DL-2). Additional thirty patients were scanned without body-weight-adapted dose reduction (DL-1-SD). Three readers evaluated subjective image quality regarding image quality and assessment of several anatomic regions. For objective image quality, signal-to-noise ratio and contrast-to-noise ratio were calculated for temporalis and masseteric muscle and the floor of the mouth. Radiation dose was evaluated by comparing the computed tomography dose index (CTDIvol) values.
Results
Deep learning-based reconstruction algorithms significantly improved subjective image quality (diagnostic acceptability: DL‑1 vs AIDR OR of 25.16 [6.30;38.85], p < 0.001 and DL‑2 vs AIDR 720.15 [410.14;> 999.99], p < 0.001). Although higher doses (DL-1-SD) resulted in significantly enhanced image quality, DL‑2 demonstrated significant superiority over all other techniques across all defined parameters (p < 0.001). Similar results were demonstrated for objective image quality, e.g. image noise (DL‑1 vs AIDR OR of 19.0 [11.56;31.24], p < 0.001 and DL‑2 vs AIDR > 999.9 [825.81;> 999.99], p < 0.001). Using weight-adapted kV reduction, very low radiation doses could be achieved (CTDIvol: 7.4 ± 4.2 mGy).
Conclusion
AI-based reconstruction algorithms in ultra-high resolution head and neck imaging provide excellent image quality while achieving very low radiation exposure.