Please use this identifier to cite or link to this item: http://doi.org/10.25358/openscience-9504
Full metadata record
DC FieldValueLanguage
dc.contributor.authorWittenmeier, Eva-
dc.contributor.authorSchmidtmann, Irene-
dc.contributor.authorHeese, Pascal-
dc.contributor.authorMüller, Pascal-
dc.contributor.authorDidion, Nicole-
dc.contributor.authorKriege, Marc-
dc.contributor.authorKomorek, Yannick-
dc.contributor.authorPirlich, Nina-
dc.date.accessioned2023-09-29T07:45:27Z-
dc.date.available2023-09-29T07:45:27Z-
dc.date.issued2023-
dc.identifier.urihttps://openscience.ub.uni-mainz.de/handle/20.500.12030/9522-
dc.description.abstractIntroduction Continuously assessing the oxygenation levels of patients to detect and prevent hypoxemia can be advantageous for safe anesthesia, especially in neonates and small infants. The oxygen reserve index (ORI) is a new parameter that can assess oxygenation through a relationship with arterial oxygen partial pressure (PaO2). The aim of this study was to examine whether the ORI provides a clinically relevant warning time for an impending SpO2 (pulse oximetry hemoglobin saturation) reduction in neonates and small infants. Methods ORI and SpO2 were measured continuously in infants aged <2 years during general anesthesia. The warning time and sensitivity of different ORI alarms for detecting impending SpO2 decrease were calculated. Subsequently, the agreement of the ORI and PaO2 with blood gas analyses was assessed. Results The ORI of 100 small infants and neonates with a median age of 9 months (min–max, 0–21 months) and weight of 8.35 kg (min–max, 2–13 kg) were measured. For the ORI/PaO2 correlation, 54 blood gas analyses were performed. The warning time and sensitivity of the preset ORI alarm during the entire duration of anesthesia were 84 s (25th–75th percentile, 56–102 s) and 55% (95% CI 52%–58%), and those during anesthesia induction were 63 s (40–82 s) and 56% (44%–68%), respectively. The positive predictive value of the preset ORI alarm were 18% (95% CI 17%–20%; entire duration of anesthesia) and 27% (95% CI 21%–35%; during anesthesia induction). The agreement of PaO2 intervals with the ORI intervals was poor, with a kappa of 0.00 (95% CI = [−0.18; 0.18]). The weight (p = .0129) and height (p = .0376) of the infants and neonates were correlated to the correct classification of the PaO2 interval with the ORI interval. Conclusions The ORI provided an early warning time for detecting an impending SpO2 decrease in small infants and neonates in the defined interval in this study. However, the sensitivity of ORI to forewarn a SpO2 decrease and the agreement of the ORI with PaO2 intervals in this real-life scenario were too poor to recommend the ORI as a useful early warning indicator for this age group.en_GB
dc.description.sponsorshipDeutsche Forschungsgemeinschaft (DFG)|491381577|Open-Access-Publikationskosten 2022–2024 Universität Mainz - Universitätsmedizin-
dc.language.isoengde
dc.rightsCC BY*
dc.rights.urihttps://creativecommons.org/licenses/by/4.0/*
dc.subject.ddc610 Medizinde_DE
dc.subject.ddc610 Medical sciencesen_GB
dc.titleEarly warning for SpO2 decrease by the oxygen reserve index in neonates and small infantsen_GB
dc.typeZeitschriftenaufsatzde
dc.identifier.doihttp://doi.org/10.25358/openscience-9504-
jgu.type.dinitypearticleen_GB
jgu.type.versionPublished versionde
jgu.type.resourceTextde
jgu.organisation.departmentFB 04 Medizinde
jgu.organisation.number2700-
jgu.organisation.nameJohannes Gutenberg-Universität Mainz-
jgu.rights.accessrightsopenAccess-
jgu.journal.titlePediatric anesthesiade
jgu.journal.volumeVersion of Record (VoR)de
jgu.publisher.year2023-
jgu.publisher.nameWiley-Blackwellde
jgu.publisher.placeOxford u.ade
jgu.publisher.issn1460-9592de
jgu.organisation.placeMainz-
jgu.subject.ddccode610de
jgu.publisher.doi10.1111/pan.14743de
jgu.organisation.rorhttps://ror.org/023b0x485-
Appears in collections:DFG-491381577-H

Files in This Item:
  File Description SizeFormat
Thumbnail
early_warning_for_spo2_decrea-20230901105114311.pdf877.79 kBAdobe PDFView/Open