Bitte benutzen Sie diese Kennung, um auf die Ressource zu verweisen: http://doi.org/10.25358/openscience-9504
Autoren: Wittenmeier, Eva
Schmidtmann, Irene
Heese, Pascal
Müller, Pascal
Didion, Nicole
Kriege, Marc
Komorek, Yannick
Pirlich, Nina
Titel: Early warning for SpO2 decrease by the oxygen reserve index in neonates and small infants
Online-Publikationsdatum: 29-Sep-2023
Erscheinungsdatum: 2023
Sprache des Dokuments: Englisch
Zusammenfassung/Abstract: Introduction Continuously assessing the oxygenation levels of patients to detect and prevent hypoxemia can be advantageous for safe anesthesia, especially in neonates and small infants. The oxygen reserve index (ORI) is a new parameter that can assess oxygenation through a relationship with arterial oxygen partial pressure (PaO2). The aim of this study was to examine whether the ORI provides a clinically relevant warning time for an impending SpO2 (pulse oximetry hemoglobin saturation) reduction in neonates and small infants. Methods ORI and SpO2 were measured continuously in infants aged <2 years during general anesthesia. The warning time and sensitivity of different ORI alarms for detecting impending SpO2 decrease were calculated. Subsequently, the agreement of the ORI and PaO2 with blood gas analyses was assessed. Results The ORI of 100 small infants and neonates with a median age of 9 months (min–max, 0–21 months) and weight of 8.35 kg (min–max, 2–13 kg) were measured. For the ORI/PaO2 correlation, 54 blood gas analyses were performed. The warning time and sensitivity of the preset ORI alarm during the entire duration of anesthesia were 84 s (25th–75th percentile, 56–102 s) and 55% (95% CI 52%–58%), and those during anesthesia induction were 63 s (40–82 s) and 56% (44%–68%), respectively. The positive predictive value of the preset ORI alarm were 18% (95% CI 17%–20%; entire duration of anesthesia) and 27% (95% CI 21%–35%; during anesthesia induction). The agreement of PaO2 intervals with the ORI intervals was poor, with a kappa of 0.00 (95% CI = [−0.18; 0.18]). The weight (p = .0129) and height (p = .0376) of the infants and neonates were correlated to the correct classification of the PaO2 interval with the ORI interval. Conclusions The ORI provided an early warning time for detecting an impending SpO2 decrease in small infants and neonates in the defined interval in this study. However, the sensitivity of ORI to forewarn a SpO2 decrease and the agreement of the ORI with PaO2 intervals in this real-life scenario were too poor to recommend the ORI as a useful early warning indicator for this age group.
DDC-Sachgruppe: 610 Medizin
610 Medical sciences
Veröffentlichende Institution: Johannes Gutenberg-Universität Mainz
Organisationseinheit: FB 04 Medizin
Veröffentlichungsort: Mainz
ROR: https://ror.org/023b0x485
DOI: http://doi.org/10.25358/openscience-9504
Version: Published version
Publikationstyp: Zeitschriftenaufsatz
Nutzungsrechte: CC BY
Informationen zu den Nutzungsrechten: https://creativecommons.org/licenses/by/4.0/
Zeitschrift: Pediatric anesthesia
Version of Record (VoR)
Verlag: Wiley-Blackwell
Verlagsort: Oxford u.a
Erscheinungsdatum: 2023
ISSN: 1460-9592
DOI der Originalveröffentlichung: 10.1111/pan.14743
Enthalten in den Sammlungen:DFG-491381577-H

Dateien zu dieser Ressource:
  Datei Beschreibung GrößeFormat
Miniaturbild
early_warning_for_spo2_decrea-20230901105114311.pdf877.79 kBAdobe PDFÖffnen/Anzeigen