Bitte benutzen Sie diese Kennung, um auf die Ressource zu verweisen:
http://doi.org/10.25358/openscience-8277
Autoren: | Waldeck, Stephan Schmidt, Sandra Falck, Christian von Chapot, René Brockmann, Marc Overhoff, Daniel |
Titel: | New hybrid multiplanar cone beam computed tomography-laser-fluoroscopic-guided approach in cochlear implant surgery |
Online-Publikationsdatum: | 20-Dez-2022 |
Erscheinungsdatum: | 2022 |
Sprache des Dokuments: | Englisch |
Zusammenfassung/Abstract: | Purpose Cochlea implant surgery with proper positioning of the cochlear electrode can be challenging. Intraoperative real-time hybrid laser-fluoroscopic-guided navigation based on a multiplanar cone beam computed tomography (CBCT) dataset opens up the opportunity to immediate radiological control of primary electrode misalignments and offering new insights into the cochlea electrode insertion routes and favorable cochlear implant-insertion angle. Methods In this retrospective study, 50 cases (29 males, 18 females) of conventional electrode implantation (without intraoperative image control; group A) and nine cases (7 males, 2 females) of CBCT-laser-fluoroscopic-guided surgery (group B) were included in the present study. CBCT-laser-guided surgery under real-time fluoroscopic control was conducted using an intraoperative C-arm CBCT. All patients received preoperative cross-sectional imaging (CT and MRI), in which cochlear malformation could be excluded. Postoperatively, we looked for electrode misplacements. Results In group A, electrode misalignment was detected postoperatively in 14 of 50 cases (28.0%). In group B, primary electrode misalignment was detected intraoperatively in two patients (22.2%). In both patients, the misalignments were corrected in the same session. The comparison of cochlear insertion angles showed significant differences. Group A: 47.5 ± 2.6° (actual conventional surgery) vs 17.6 ± 2.8° (theoretical CBCT-laser-fluoroscopic-guided surgery) P < 0.001. Group A vs group B: 47.5 ± 2.6° (actual conventional surgery; Group A) vs 17.9 ± 2.5° (actual CBCT-laser-fluoroscopic-guided surgery; Group B) P < 0.001. Conclusion We consider that an intraoperative hybrid CBCT-laser-fluoroscopic-controlled approach in cochlear implant surgery using a C-arm CT can be beneficial, because electrode misalignments can be reduced and if it does occur, remedied in the same surgical session. |
DDC-Sachgruppe: | 610 Medizin 610 Medical sciences |
Veröffentlichende Institution: | Johannes Gutenberg-Universität Mainz |
Organisationseinheit: | FB 04 Medizin |
Veröffentlichungsort: | Mainz |
ROR: | https://ror.org/023b0x485 |
DOI: | http://doi.org/10.25358/openscience-8277 |
Version: | Published version |
Publikationstyp: | Zeitschriftenaufsatz |
Nutzungsrechte: | CC BY |
Informationen zu den Nutzungsrechten: | https://creativecommons.org/licenses/by/4.0/ |
Zeitschrift: | International journal of computer assisted radiology and surgery 17 |
Seitenzahl oder Artikelnummer: | 1837 1843 |
Verlag: | Springer |
Verlagsort: | Berlin u.a. |
Erscheinungsdatum: | 2022 |
ISSN: | 1861-6429 |
DOI der Originalveröffentlichung: | 10.1007/s11548-022-02703-2 |
Enthalten in den Sammlungen: | DFG-491381577-H |
Dateien zu dieser Ressource:
Datei | Beschreibung | Größe | Format | ||
---|---|---|---|---|---|
new_hybrid_multiplanar_cone_b-20221114110901631.pdf | 714.19 kB | Adobe PDF | Öffnen/Anzeigen |